Abstract
To utilize the transmit radiofrequency (RF) field in MRI as a power source, near or within the field of view but without affecting image quality or safety. Power harvesting is performed by RF induction in a resonant coil. Resulting RF field distortion in the subject is canceled by a selective shield that couples to the harvester while being transparent to the RF transmitter. Such shielding is designed with the help of electromagnetic simulation. A shielded harvester of 3 cm diameter is implemented, assessed on the bench, and tested in a 3T MRI system, recording power yield during typical scans. The concept of selective shielding is confirmed by simulation. Bench tests show effective power harvesting in the presence of the shield. In the MRI system, it is confirmed that selective shielding virtually eliminates RF perturbation. In scans with the harvester immediately adjacent to a phantom, up to 100 mW of average power are harvested without affecting image quality. Selective shielding enables stealthy RF harvesting which can be used to supply wireless power to on-body devices during MRI.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.