Abstract

In this paper we evaluate methods for standardization of Raman spectra that are required to improve spectral correlation computations between spectra measured on different instruments. Five commercially-available 785 nm Raman spectrometers from different vendors were included in the study. These spectrometers have diverse specifications and performance levels and range in size from laboratory-based instruments to field-deployable portable and handheld platforms. Since each Raman spectrometer has different characteristics, spectra obtained on one instrument cannot readily be compared to a library acquired on a different instrument without performing various types of spectral corrections (standardization). We outline a procedure that combines previously established Raman shift and intensity correction protocols with a resolution matching step to facilitate the comparison of a centralized master library with spectra acquired on different geographically distributed Raman spectrometers. The standardization procedure is effective in reducing the inherent instrument-to-instrument variability so that spectra from different spectrometers can be compared and reliable results obtained using library-based spectral correlation methods. The findings have important implications for the ability to transfer Raman spectral libraries between instruments.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call