Abstract

Two methods were used for the standardization of the pure beta-emitter 89Sr solution received in the framework of the 2000 BIPM comparison of activity measurements. The first method was by scintillation counting using the triple to double coincidence ratio (TDCR) technique. The existing three detector-system was improved by the inclusion of a MAC-3 module produced by LNHB-Saclay. This module contains the circuits for the coincidence, dead time and gate functions. This was the first use of this equipment in an international comparison. The DETECSZ program (LNHB) was used to obtain the activity concentration, at the reference time, as 26.09±0.21 kBq g −1. The second method was the efficiency tracer technique, using a 60Co standard solution and a 4πβ–γ-coincidence system. The radioactivity concentration was calculated from a general extrapolation curve, by using data from several sources. The value was 26.58±0.28 kBq g −1 on the reference date 01.10.2000, 00h UTC. The two results are in agreement within the limits of their combined uncertainties.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.