Abstract

For the estimation of the tritium activity in environmental samples, the triple to double coincidence ratio (TDCR) method is standardized for the liquid scintillation counter (LSC). Variation of TDCR with tritium activities, optimization of the region of interest for low-level measurement of tritium, the effect of different quenching agents on the TDCR values, and application of TDCR method for the quantification of ionization quench factor (kB) are discussed in this paper. Relative standard deviation in TDCR values at tritium activity levels of 15–140 Bq/L is found to be less than 1%. The the quench curve equation does not depend on the type of quenching agents.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.