Abstract

A special stabilization system has been constructed to control the overall gain of photomultiplier tubes in a liquid scintillation counter. The system uses GaAsP light emitting diodes in the measurement chamber to produce separate reference light pulses for each photomultiplier tube. The system has provided a significant improvement in both short and long term stability and in operational reliability.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.