Abstract

A quality by design (QbD) based high-resolution HPLC method is described for determination of impurities in apixaban (APX) in the tablet dosage form. Employing a simple and stability-indicating HPLC method, nine known impurities were quantified with good peak resolution. Mobile phase A (MP-A) was prepared with buffer and acetonitrile 90:10v/v, while mobile phase B (MP-B) contained water and acetonitrile 10:90 v/v. The gradient program was 0 min, MP-A 75%, B 25%; 20 min, MP-A 65%, B 35%; 30 min, MP-A 40%, B 60%; 40min, MP-A 40%, B 60%; 42 min, MP-A 75%, B 25%; and 50 min, MP-A 75%, B 25%. The chromatographic separation was achieved using a Zorbax RX C18 250 × 4.6 mm column, 5 μm (1.0 ml min-1 , 280 nm, 50 μl) and a column temperature of 40°C. Several separation studies were carried out using design of experiments to optimize the method. Validation results confirm the applicability of the developed method for quality analysis and stability studies of the regular product on the manufacturing stream.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call