Abstract

A simple, accurate, rapid, and stability-indicating RP-HPLC method for a combination of tenofovir disoproxil fumarate, emtricitabine, and rilpivirine has been developed and subsequently validated in commercial tablets. The proposed HPLC method utilizes Phenomenex Gemini C18 column (150 mm × 4.6 mm i.d., 5 µm) and mobile phase consisting of MeCN, potassium dihydrogen phosphate buffer (20 mM, pH 3.3), and triethylamine 58.72 : 41.23 : 0.05 (v/v) at a flow rate of 1.7 mL/min. Quantitation was achieved with UV detection at 270 nm. The method was validated in terms of accuracy, precision, linearity, limits of detection, limits of quantitation, and robustness. This optimized method has been successively applied to pharmaceutical formulation and no interference from the tablet excipients was found. TDF, EMT, and RPV and their combination drug product were subjected to acid, base, neutral hydrolysis, oxidation, dry heat, and photolytic stress conditions and the stressed samples were analyzed by the proposed method. As the proposed LC method could effectively separate the drugs from its degradation products, it can be employed as stability-indicating method for the determination of instability of these drugs in bulk and commercial tablets.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.