Abstract

A simple, accurate reverse phase high performance liquid chromatographic method has been developed for the simultaneous estimation of lamivudine, tenofovir disoproxil fumarate and efavirenz in bulk and pharmaceutical formulations. The analytical method development was carried on Agilent make HPLC instrument using RP - C18 column. The mobile phase employed for the estimation is phosphate Buffer pH 4.0 :acetonitrile adjusted to pH 4.0 with glacial acetic acid which was pumped at a flow rate of 1.0 mL min-1 in the ratio of 42:58 v/v. the eluents were monitored at 260 nm. Linearity was obtained in the concentration range of 20-100 μg/mL of lamivudine, tenofovir disoproxil fumarate and 100-500 μg/mL efavirenz. Degradation studies shows that all the three drugs were not degraded under acidic, alkaline, thermal and photolytic conditions.The method was statistically validated and RSD was found to be within limits. Due to its simplicity, rapidness, high precision and accuracy, the proposed HPLC method can be applied for determining lamivudine, tenofovir disoproxil fumarate and efavirenz in bulk and in pharmaceutical dosage form.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.