Abstract
The paper presents experimentally obtained results on the long-term stability tests of the solid electrolyte amperometric sensor output signal during the operation in the O2 + N2 gas mixtures. These data prove the stability and reproducibility of the output signal when measuring the oxygen concentration in air for 8000 hours. The output signal variation during the tests did not exceed ± 2 %. We performed four heating / cooling cycles of different durations, which did not influence either sensor integrity or operation characteristics. The structure of the solid electrolyte sensor and the solid electrolyte/electrode interfacial layer remained unchanged during the tests. Dynamic characteristics of the sensor, including the response time, were stable.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.