Abstract

Stability analysis and design of an automatic frequency control (AFC) system for in vivo continuous-wave EPR spectroscopy is described. The open-loop function of the feedback control system for the AFC was derived and the stability of the feedback loop systematically examined. A stability analysis of the system is demonstrated and a systematic design procedure is proposed. The design is started from the required system specifications (phase margin, steady-state error, and system bandwidth) and clear guidelines for designing an AFC system are given. A case study of the design is presented based on the specific needs of in vivo EPR measurements. A phase margin of 53 degrees, a steady-state error of 1.6%, and a system bandwidth of up to 1.8 kHz were obtained in the designed AFC system. The system specifications defined in advance are satisfied in this case study.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.