Abstract

One normally assumes that the analytical measurement process does not introduce spurious compounds. Capillary electrophoresis is a separation method frequently used for small-volume biological measurements. We demonstrate the potential for creating new peaks in a capillary electropherogram when using electrokinetic injections and illustrate the potential deleterious effects with biological samples involving serotonin and nitric oxide measurements. Specifically, when measuring the serotonin content from individual neurons using electrokinetic injections from 360 nL stainless steel vials, we detect a new peak that we identify as a serotonin dimer. We do not observe this peak when using hydrodynamic injections.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.