Abstract

A sensitive, precise and reliable flow-batch method for the determination of aluminium (Al) was developed using a sequential injection-monosegmented flow system incorporating a mixing chamber unit. Eriochrome cyanine R (ECR) was used as a chromogenic reagent in the presence of N,N-dodecyltrimethylammonium bromide (DTAB). The Al-ECR complex at pH 6 gave a maximum absorption at 584nm. In-line single standard calibration and a standard addition procedure were developed employing the monosegmented flow technique. Under the optimum conditions, a linear calibration graph in the range of 0.0075–0.625mgL−1 Al was obtained with limits of detection and quantitation of 0.0020 and 0.0070mgL−1, respectively. Relative standard deviations were 0.8 and 1.3% for 0.010 and 0.025mgL−1 Al (n=11), respectively. A sample throughput of 24h−1 using an in-line standard calibration approach and 6h−1 using four standard addition levels was achieved. The developed system was successfully applied to water samples and beverage samples. The results agreed well with those obtained from the ICP-AES method. Good recoveries between 85 and 104% were obtained.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.