Abstract

Abstract Serum samples are deproteinized with an equal volume of acetone and 20 μl of the filtrate is added to 1 ml of 2.5 × 10−4 M KOH and 1.0 × 10−4 M 1, 8-dihydroxyanthraquinone in 90% (V/V) acetone. Single point calibration with spiked serum at 1.5 mmol/l is used with unspiked serum as blank. Fluorescence emission is measured at 618 nm, with excitation at 410 nm. Samples analyzed by the described method and by atomic absorption spectrometry showed a correlation coefficient of 0.985. The relative responses for lithium in water, serum eletrolyte matrix and serum were 100, 57, and 44, respectively. The Li response in serum relative to that for Na, K, Ca, and Mg was 480, 600, 33, and 27, respectvely.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.