Abstract
A method is described of measuring absolute spectral response for Au and CsI transmission photocathodes in soft X-ray streak camera, which is of great importance for the inertial confinement fusion (ICF) diagnostics. Transmission photocathode is conventionally employed as photo-to-electron conversion accessories. To derive quantity information of X-ray spectra, the absolute response of photocathode must be calibrated in a range of interest. Here Au and CsI transmission photocathodes with slits are calibrated respectively on Beijing Synchrotron Radiation Facility (BSRF), in a photon energy range of 60 eV5500 eV. This method has an uncertainty less than 10% and good feasibility. Calibration results are in good agreement with the calculation results obtained from the Henke's photon emission model, with CH substrate effect revised.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.