Abstract
The spectral emissivity and transmissivity of zinc sulphide (ZnS) infrared windows in the spectral region from 2 to 12 μm and temperature range from 20 to 700°C is measured by a facility built at the Harbin Institute of Technology (HIT). The facility is based on the integrating-sphere reflectometry. Measurements have been performed on two samples made of ZnS. The results measured at 20°C are in good agreement with those obtained by the method of radiant energy comparison using a Fourier transform infrared spectrometer. Emissivity measurements performed with this facility present an uncertainty of 5.5% (cover factor=2 ).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.