Abstract
Herein, a method was established for the speciation of six selenium species by high performance liquid chromatography-inductively coupled plasma-tandem mass spectrometry (HPLC-ICP-MS/MS). The factors affecting separation were carefully investigated, including ionic strength, pH, and methanol content. Six species of selenium could be completely separated within 20 min, under the mobile phase of 25 mM citric acid in pH = 4.0 containing 2% methanol. The detection limits of selenite (Se(IV)), selenate (Se(VI)), selenomethionine (SeMet), selenocystine (SeCys2), methylselenocysteine (MeSeCys), and selenoethionine (SeEt) were 0.04, 0.02, 0.05, 0.02, 0.03, and 0.15 ng mL−1, respectively. To verify the practicality of this method, the analysis of selenium-enriched foods such as selenium-enriched spring water, selenium-enriched salts, and selenium-enriched tea were conducted, and recovery of 93.7–105% was achieved with RSD < 5%, revealing the high practical utility of the proposed method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.