Abstract

A pretreatment method of dispersive solid-phase extraction (DSPE) along with back-extraction followed by CE-UV detector was developed for the determination of mercury species in water samples. Sulfhydryl-functionalized SiO2 microspheres (SiO2 -SH) were synthesized and used as DSPE adsorbents for selective extraction and enrichment of three organic mercury species namely ethylmercury (EtHg), methylmercury (MeHg), and phenylmercury (PhHg), along with L-cysteine (L-cys) containing hydrochloric acid as back-extraction solvent. Several main extraction parameters were systematically investigated including sample pH, amount of adsorbent, extraction and back-extraction time, volume of eluent, and concentration of hydrochloric acid. Under optimal conditions, good linearity was achieved with correlation coefficients over 0.9990, in the range of 4-200μg/L for EtHg, and 2-200μg/L for MeHg and PhHg. The LODs were obtained of 1.07, 0.34, and 0.24μg/L for EtHg, MeHg, and PhHg, respectively, as well as the LOQs were 3.57, 1.13, and 0.79μg/L, respectively, with enrichment factors ranging from 109 to 184. Recoveries were attained with tap and lake water samples in a range of 62.3-107.2%, with relative standard deviations of 3.5-10.1%. The results proved that the method of SiO2 -SH based DSPE coupled with CE-UV was a simple, rapid, cost-effective, and eco-friendly alternative for the determination of mercury species in water samples.