Abstract

The plasma shielding effect is one of the major weaknesses of laser-induced breakdown spectroscopy (LIBS) as it causes non-linearity in signal strength. Although LIBS is typically carried out in constant laser energy, this non-linearity causes a reduction in sensitivity. In this work, we systematically examine laser-induced plasma, formed by two different excitation source modes, i.e. single pulse (SP)-excitation and single-beam-splitting double-pulse (SBS-DP)-excitation over Zr-2.5% Nb alloy. The two most important plasma parameters influencing the emission line intensity, plasma temperature (T e) and electron density (N e) were studied and compared for both modes of laser excitation. Comparison of the results conclusively demonstrates that due to the splitting of the laser energy in the SBS-DP mode, the plasma shielding effect is significantly reduced. The reduced plasma shielding translates to an increased laser–sample coupling under SBS-DP mode. Temporal imaging of the total intensity of the laser-induced plasma in both excitation modes was also studied. The study shows how the plasma shielding effect can be reduced to improve the analytical quality of the LIBS methodology.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.