Abstract

The quantification of lithium in electron microscopes by energy-dispersive X-ray spectroscopy (EDS) is highly challenging due to the low energy of the characteristic X-rays. Only special windowless detectors can detect Li, yet the sensitivity is suboptimal (i.e., ~20 wt. %). We introduce the use of quantitative backscattered electron microscopy in combination with measurements using a standard EDS sensor to achieve spatially resolved quantification of Li. The method is demonstrated using two Mg-Li alloys with 4 and 20 wt. % Li and good accuracy with a deviation of only ~1.2 wt. % and ~2.5 wt. %, respectively, is achieved.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.