Abstract
Objective To identify the spatial distribution of stray radiation from mobile CT head scanning for the purpose of radiation protection. Methods The head series of CareTom mobile CT were scanned and the radiation dose was measured using TLD (LiF: Mg, Cu, P). The isodose maps of radiation dose field were plotted using Matlab software. Results Radiation dose in the front of the mobile CT was slightly higher than that in the back. The maximum value of 0.255 mGy was found to be at 0.5 m from the scanning hole center. Conclusions The stray radiation dose from mobile CT head scanning was relatively low. However in order to avoid the damage to the operators and other medical workers from long-term low dose exposure, it should keep 2 m away from mobile CT, beside or behind, when in operation. Key words: Mobile CT; Head scanning; Radiation dose field
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.