Abstract
NaI(Tl) scintillation detectors are widely used in measurements of environmental radioactivity, low level radioactive waste, prompt gamma ray neutron activation analysis, some nuclear physics experiments, geology, etc. This paper reports the results of the energy calibration, resolution and full energy peak efficiency and total efficiency calculations for a 3"x3" NaI(Tl) detector (Canberra Inc.) that are essential to specify the quality of the results of gamma spectrometry measurements. To investigate the effects of the gamma ray energy and the distance from source to detector center on detector efficiency, measurements were performed at different six axial distances for four point sources (Am-241, Cs-137, Co-60 and Na-22). Thus, fundamental data for further works with this detector system were obtained.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.