Abstract

A model is presented describing the effects of dead time and mass bias correction factor uncertainties, flicker noise, and counting statistics on isotope ratio measurement precision using inductively coupled plasma mass spectrometry (ICPMS) with a single collector. Noise spectral analysis is exploited to enable estimation of the flicker noise parameters. For the instrument used, the flicker noise component exhibited a fairly weak frequency (t) dependence (is proportional to f -0.33+/-0.12), but was directly proportional to the total number of counts, Q. As white noise, determined by counting statistics, is given by Q0.5, the isotope ratio measurement uncertainties will actually cease to improve when Q exceeds a certain threshold. This would suggest that flicker noise could become the limiting factor for the precision with which isotope ratios can be determined by ICPMS. However, under most experimental conditions, uncertainties associated with mass discrimination and dead time correction factors are decisive. For ratios up to approximately 22 (115In/113In), optimum major isotope count rates are generally below 0.3 MHz, for which precision in the mass discrimination factor is limiting. The model derived could be used as a starting point for determining optimum conditions and understanding the limitations of single-collector ICPMS for precise isotope ratio measurements.