Abstract

The determination of manganese(II) in tap water and effluent streams, using a solid-phase reactor incorporated into a sequential injection system was investigated. Mn2+-ions in samples injected into a carrier stream, were oxidised by solid lead(IV) dioxide suspended on silica gel beads to form MnO4- -ions which were detected spectrophotometrically at 526 nm. The linear range of the system is from 1 to 7 mg L(-1) with a detection limit of 0.62 mg L(-1). The proposed system is suitable for the determination of manganese(II) in tap and effluent streams at a rate of approximately 50 samples per hour with a relative standard deviation of better than 3%. Statistical comparison between the proposed sequential injection system and a standard ICP method revealed that there is no significant difference between the two methods at the 95% confidence level for effluent streams and at 99.9% for tap water.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.