Abstract
The determination of manganese(II) in natural water and effluent streams, using a solid-phase reactor incorporated into a flow-injection system, was investigated. Mn 2+-ions in samples injected into a carrier stream, were oxidised by solid lead(IV) dioxide suspended on silica gel beads to form MnO 4 −-ions which were detected spectrophotometrically at 526 nm. The linear range of the system is between 1 to 5 mg l −1 with a detection limit of 0.56 mg l −1. The proposed system is suitable for the determination of manganese(II) in natural water and effluent streams with a relative standard deviation of better than 1.8%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.