Abstract

The aim of this study was to develop a methodology for the analysis of the insecticide fenitrothion and its two main environmental metabolites, fenitrooxon and 3-methyl-4-nitrophenol. For this purpose, a solid-phase microextraction (SPME) method coupled to high performance liquid chromatography (LC) was optimized. Two on-line detectors, diode array (DAD) and direct current amperometrical (DCAD) were used in order to determine sensitivity and selectivity. The effects of the extraction parameters, including exposure and desorption time, pH, temperature, salt concentration and desorption mode on the extraction efficiency were studied. A satisfactory reproducibility for extractions from samples at 20 ppb-level with RSD < 12.5% ( n = 10) was obtained. The calibration graphs were linear in the range of 10–1000 μg l −1 and detection limits for the target compounds were between 1.2 and 11.8 μg l −1 depending on which detector was used. The method was applied for determining fenitrothion and both its metabolites in river waters which run through forest areas near to aerial application of the pesticide.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.