A method for determining urine porphyrins by HPLC is described. In the preliminary step, porphyrins are purified in high yields and concentrated by low pressure reverse-phase chromatography on C18 (octadecylsilane bonded silica) cartridge. Porphyrins are stable for 10 days after adsorption on C18 cartridge. The separation of porphyrin esters is performed on an aminopropyl-bonded silica column with an eluting system containing n-heptane and ethyl acetate. The system enables rapid isocratic separation of porphyrin methyl esters with high selectivity. The simplicity and reproducibility of the whole procedure allows its application to the routine analysis of urinary porphyrins in the clinical laboratory.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.