Abstract
The dosimetric properties of the SOL-GEL α-Al2O3 detectors with several different concentrations of impurities were studied in beta radiation beams. For the characterization of the detectors, the TL and OSL techniques were used. A TL spectrum was acquired. The detectors showed adequate OSL and TL dosimetric characteristics, such as good reproducibility, a glow curve with the peak at ~230 °C at 5 °C/s of heating rate, adequate OSL decay curve, good linearity of response and high phosphor sensitivity. A thermal treatment study was also performed. Moreover, the calibration factors and the lower detection limits were determined for beta radiation. These characteristics indicate the suitability of SOL-GEL α-Al2O3 detectors for the establishment of a transfer system or alternative/complementary method for beta radiation dosimetry.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.