Abstract

The OSL and the TL techniques are used in several fields of radiation dosimetry. The dosimeters are affected by a signal loss in the time between the irradiation and readout (fading). This undesirable characteristic can cause an underestimation of the irradiation dose. The fading estimation of SOL-GEL α-Al2O3 with several different concentrations of impurities is reported. The thermal fading and the light-induced fading of the SOL-GEL α-Al2O3 detectors were estimated. Furthermore, the exponential equations used to interpolate the experimental results were presented.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.