Abstract

A small computer data system was constructed and evaluated for use with a portable quadrupole mass filter. The system is composed of two conventional audio tape decks, one interfaced to the portable mass filter and the other to a small shared computer. In operation, a series of mass spectra is recorded on magnetic tape. Next, the tape is transferred to the audio tape deck interfaced to the computer. The computer adds the spectra together and performs appropriate spectrum stripping and arithmetic routines. By reexamining hydrogen-deuterium self-exchange in water, we show that the mass filter-recorder-computer system gives results that are as accurate and precise as results obtained from other mass spectrometric methods.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.