Abstract

A simple and robust method for the determination of arsenic in sediments by hydride generation atomic absorption spectrometry (HG-AAS) using slurry sampling is proposed. Aqua regia (5% v/v) and hydrofluoric acid (1% v/v) were mixed with an aliquot of the sediment with particle size of 50 µm in an ultrasonic bath for 30 min. The slurry was allowed to stand for 48 h for an effective extraction of the analyte to the aqueous phase. The final volume of the slurry was made up with HCl 1 mol L-1. A NaBH4 solution stabilized with NaOH was used as reducing agent. The arsine was transported to the quartz T-tube heated at 900 °C for detection. Five certified reference materials of sediment were analyzed and the obtained concentrations were in good agreement with the certified values, according to the t-test for a 95% confidence level with RSD (relative standard deviation) lower than 13%. The limit of detection (LOD) in the sample was 0.6 µg g-1. The proposed procedure for arsenic determination in sediment is precise, accurate and adequate for the environmental samples routine analyses.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.