Abstract

Conversion of slow scan signals with a continuously updated image display or video recording at TV rate is one of the functions available using a digital video framestore system in conjunction with a scanning electron microscope (SEM, STEM, AEM etc.) or other types of electron probe equipment e.g. EPMA, AES etc. Acceptable inputs are standard TV format; digital sampling of a range of analogue slow scans or directly coupled digital scans.Routines for noise reduction by spatially resolved time averaging during acquisition are under software control with direct operator involvement in selecting the appropriate conditions via a soft-key redesignation of the VDU numeric pad functions (Fig.l) and within the possibilities of the hardware; making TV rate more useable e.g. to reduce specimen charging and as an essential prerequisite to more sophisticated signal/image processing. Both running average (RVP, 8-bit) and straight integration (16-bit) modes are possible with the Microconsultants Intellect systems in use in our laboratory.