Abstract
We report, here, a biosensor based on flow injection of the recognition element (glucose oxidase), applied on small volume (30 μl) whole blood samples. This SIRE (sensors with injectable recognition elements) technology based biosensor gave a linear response in the range of 0-12 mM glucose. A repeatability of 5.1% (n=11) and a reproducibility of 8.0 % (n=4) was observed when using glucose standard solutions. A glucose tolerance monitoring, after a peroral intake, was performed. Our clinical analysis, shown in this report, illustrates that the results obtained by the SIRE biosensor are comparable to results obtained from a commercial device (Medisense, USA). The advantages of the presented biosensor, concerning multi analyte analysis and background signal control, are focused upon. These features makes the biosensor especially suitable for real time bed side monitoring at hospitals.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.