Abstract

The purpose of this work was to perform an independent and National Institute of Standards and Technology-traceable activity measurement of 90Y SIR-Spheres (Sirtex). γ-spectroscopic measurements of the 90Y internal pair production decay mode were made using a high-purity germanium detector. Methods: Measured annihilation radiation detection rates were corrected for radioactive decay during acquisition, dead time, source attenuation, and source geometry effects. Detection efficiency was determined by 2 independent and National Institute of Standards and Technology-traceable methods. Results: Measured SIR-Spheres vials (n = 5) contained more activity than specified by the manufacturer calibration; on average, the ratio of measured activity to calibrated was 1.233 ± 0.030. Activity measurements made using 2 distinct efficiency calibration methods agreed within 1%. Conclusion: The primary SIR-Spheres activity calibration appears to be a significant underestimate of true activity.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.