A highly diluted suspension of red blood cells (hematocrit 0.01) was illuminated with an Ar or a dye laser in the wavelength range of 458-660 nm. The extinction and the angle-resolved intensity of scattered light were measured and compared with the predictions of Mie theory, the Rayleigh-Gans approximation, and the anomalous diffraction approximation. Furthermore, empirical phase functions were fitted to the measurements. The measurements were in satisfactory agreement with the predictions of Mie theory. However, better agreement was found with the anomalous diffraction model. In the Rayleigh-Gans approximation, only small-angle scattering is described appropriately. The scattering phase function of erythrocytes may be represented by the Gegenbauer kernel phase function.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.