Abstract

A single-drop microextraction (SDME) method followed by in-syringe derivatization and GC-MS determination has been developed for analysis of five parabens, including methyl, ethyl, isopropyl, n-propyl and n-butyl paraben in water samples and cosmetic products. N,O-Bis(trimethylsilyl)acetamide (BSA) was used as derivatization reagent. Derivatization reaction was performed inside the syringe barrel using 0.4 microL of BSA. Parameters that affect the derivatization yield such as temperature and time of the reaction were studied. In addition, experimental SDME parameters such as selection of organic solvent, addition of salt, extraction time and extraction temperature were investigated and optimized. The RSD of the method for aqueous samples varied from 8.1 to 13%. The LODs ranged from 0.001 (n-butyl paraben) to 0.015 (methyl paraben) microg/L, and the enrichment factors were between 23 and 150.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.