Abstract

Simultaneous two channel array proton imaging, single voxel PRESS and CSI acquisitions were demonstrated after reconfiguration and minimum hardware modification of a standard 4.7T BioSpec® spectrometer. Validation of the reconfiguration was assessed in phantoms and in a mouse brain. The modified configuration used the X channel exhibiting similar SNR performances compared to the 1H channel. The SNR gain for the two channel array coil was up to 1.3 compared to the SNR obtained with a reference surface coil. Compared to regular two element coil with quadrature combination, the SNR was improved with an additional gain of 1.3. These modifications could also be applied for any X nucleus

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.