Abstract
We present a single arm low-coherence interferometer to directly measure the physical thickness and group refractive index of optically transparent samples having flat and parallel surfaces. The optical arrangement, resembling a common-path interferometer, is more compact and stable than the usual dual-arm low-coherence interferometer. It has been used to measure samples of Herasil 102 fused silica, Schott B270 Superwhite crown glass and borosilicate cover glass. The results obtained indicate uncertainties in the third decimal place for index values and thicknesses accurate to within 2 μm.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.