Abstract

In this study, we simultaneously measured the group refractive index dispersion and thickness of fused silica using a scanning white light interferometer on a spectral range from 800 to 1050 nm. A delay error correction was performed using a He-Ne laser. The accuracy of the measured group refractive index dispersion of fused silica, when compared to the temperature-dependent Sellmeier equation, is within 4 × 10-5.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.