Abstract

Historically, analyses performed by atomic spectroscopy have been accomplished by measurement of the emission or absorption of light by an analyte species. The ability to measure both effects simultaneously may have some important applications in routine analysis (especially with respect to increasing linear dynamic response) and in the study of the atomization process. Utilization of a modified BECS background emission correction system developed by Dewalt and associates for furnace atomic absorption spectroscopy and a pulsed hollow cathode light source allows for the measurement of both analyte emission and absorption. The utilization of the Woodriff design constant temperature furnace atomizer as an atomic absorption source is well documented; its ability to produce atomic emission has been recently reported. Coupling of the furnace with the described electronics results in a system capable of producing highly accurate analyses over a wide concentration range.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.