Abstract
In this paper, a new model is presented for estimation of the blood glucose level from the measured near-infrared absorbance. The model has been developed in such a way that the regression coefficients of this linear relation have been approximated by considering only the molar absorptivity of the glucose and the obtained coefficients have been utilized to estimate the blood glucose levels from the measured absorbances. The estimation of the blood glucose concentrations by this blind approach exhibited an acceptable accuracy in comparison to the more accurate principal components regression method. The blood sample absorbances have been measured using a Fourier transform infrared device while the blood glucose levels have been determined by a commercial finger-prick glucometer device.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Similar Papers
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.