Abstract

An ion-paring HPLC–MS–MS method with positive ion mode electrospray ionization has been developed to simultaneously quantify Ziagen, carbovir monophosphate, carbovir diphosphate and carbovir triphosphate. N′, N′-Dimethylhexylamine was used as the ion-pairing agent. The presence of this ion-pairing agent allowed the retention and separation of the four compounds on a reversed-phase HPLC column as well as the detection of the nucleotides with positive ion mode electrospray ionization. The limits of detection were found to be better than 25 n M for all the analytes. Calibration curves of the analytes showed excellent linearity over the range of 25 n M to 5 μ M. The relative standard deviations and accuracies for replicate analyses of quality control samples were less than 15%. The method has been successfully applied to the analysis of these compounds in human liver cells treated with Ziagen.