Abstract

A microfluidic device for simultaneous analysis of total flavonoids and total phenolic acids is presented for the first time. The process was based on the utilization of magnetic zinc-imidazole frameworks (ZIF-4), as a highly efficient media for on-line separation of phenolic acids and flavonoids, which were subsequently detected by a sensitive chemiluminescence (CL) method. Acrylate-based polymeric microchips containing a separation column (12.5 mm length, 3 mm width and 1 mm depth) were fabricated using a 3D-printer, and magnetic ZIF-4 was fixed into the column. The high porosity of the magnetic ZIF-4 made it a suitable adsorbent, guaranteeing an effective separation. The detection limits were in the range of 0.04–0.10 µg mL−1 with relative standard deviation values of (2.19–4.38%). The system was successfully applied for the analysis of flavonoids and phenolic acids in tea and honey samples. The recovery was from 95.4 to 104.1%, indicating a good accuracy of the developed method.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call