Abstract

The quantitative analysis of solutions containing various concentrations of sodium hydroxide, sodium chloride and sodium carbonate, over the range 0–15%m/m, has been carried out by near-infrared (NIR) spectrometry. Errors, caused by a high concentration of one salt when determining small amounts of another, have been investigated. The standard errors of the computed results from NIR measurements of each ion were: hydroxide, 0.12; carbonate, 0.34; and chloride, 0.52%. The measurements were made using a liquid sample cup in a fixed-filter NIR instrument and it was concluded that this was a practical system for continuous monitoring of the sodium hydroxide concentration in a process line even when contamination from hydrochloric acid and carbon dioxide occurred.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.