Abstract

An automated method for simultaneous routine quantification of the antipsychotic drugs clozapine, olanzapine and their demethylated metabolites is described. The method included adsorption on a cyanopropyl (CPS) coated clean-up column (10 μm; 10×2.0 mm I.D.), washing off interfering serum constituents to waste, and separation on C18 ODS Hypersil reversed phase material (5 μm; 250×4.6 mm I.D.) using acetonitrile–water–tetramethylethylenediamine (37:62.6:0.4, v/v/v) adjusted to pH 6.5 with concentrated acetic acid. UV-detection was performed at 254 nm. The limit of quantification was 10–20 ng/ml. Relative day to day standard variations ranged between 4.5 and 13.5%. The method is suitable for routine monitoring of olanzapine and clozapine including their demethylated metabolites.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.