Abstract
The use of nuclear magnetic resonance (NMR) spectroscopy as a tool for determining pharmaceutical molecules in bulk drugs and their dosage forms is growing. New advancements in benchtop NMR spectrometers with cryogen-free magnets have made this technique more appealing and accessible. Herein, we developed a method using a benchtop NMR spectrometer to quantify phenytoin (PhT) and phenobarbital (PhB) in bulk and combined dosage forms. The results were compared to those obtained by high performance liquid chromatography (HPLC) as a well-characterized procedure. This method is simple, low cost, relatively fast, and non-inferior to HPLC in terms of figures of merit.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.