Abstract
A simple method was developed for the determination of trace metals (Mn, Fe and Co) in copper simultaneously using a multi-element graphite furnace atomic absorption spectrometer (Perkin-Elmer SIMAA6000) equipped with a transversely heated graphite atomizer. Copper was dissolved in nitric acid before analysis using a microwave digestion system. Pd(NO 3) 2 with the special gas (5% H 2 in Ar) was used as the chemical modifier. The accuracy of the method was confirmed by the analysis of certified reference copper (CRM#075). The interferences were removed effectively, so that a calibration curve method based on simple aqueous standards could be used. The analyzed values were in the range of 96.6% (Mn) to 113% (Co) of the certified values. The relative standard deviations for the simultaneous determination of Mn (6.5 μg/l), Fe(18 μg/l) and Co(5 μg/l) in solution from copper are 2.4–3.5% and the detection limits were 0.17 μg/g for Mn, 0.50 μg/g for Fe and 0.30 μg/g for Co.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.