Abstract

A simple method was developed for the direct and simultaneous determination of copper, chromium, aluminum, and manganese in urine using a multielement GFAAS (Perkin-Elmer SIMAA6000). Pd was used as the chemical modifier along with a special purge gas (5% H2 in Ar). A simple calibration curve method can be used (with 1:1 dilution). A standard reference material (Seronorm Trace Elements Urine) was used to find the optimal temperature program and to confirm the accuracy of the technique. The analyzed values were within 90-110% of the certified values. The relative standard deviations were 1.7, 1.5, 1.6, and 1.5% for these four elements and the detection limits were 0.08 microg L(-1) for Cu, 0.05 microg L(-1) for Cr, 0.06 microg L(-1) for Al, and 0.06 microg L(-1) for Mn. The recoveries of Cu, Cr, Al, and Mn from real urine samples were 100 +/- 5%, except for Cu (80%). The found values of Cu, Cr, Al, and Mn in a real urine sample were 14.3, 0.78, 18.9, and 0.06 microg L(-1), respectively. Scanning electron micrographs were used to investigate the physical form of Pd on the surface of the platform in the graphite furnace. Use of 5% H2 in Ar as the purge gas resulted in smaller and more uniformly distributed Pd particles (Pd particle diameters 0.4-0.6 microm using 5% H2 in Ar compared to 0.4-1.2 microm using pure Ar), increasing the effect of the Pd chemical modifier and promoting the efficiency of atomization.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.