Abstract

A liquid chromatographic system for the determination of glucose and 1-deoxyglucose in serum using an immobilized pyranose oxidase reactor and chemiluminescence detection is described. Separation was achieved on a TSK gel SAX column (2 cm × 4 mm I.D.) with an eluent of 0.1 M sodium hydroxide solution. Serum was diluted 30-fold with 0.1 M NaOH. Sample solution (50 μl) was injected into the system. The hydrogen peroxide produced was detected by measuring the chemiluminescence emitted on admixing with luminol and potassium hexacyanoferrate (III). The calibration graphs were linear from 1 to 600 μ M glucose and from 0.6 to 400 μ M 1-deoxyglucose; the detection limits for glucose and 1-deoxyglucose were 0.5 and 0.3 μ M, respectively. The sample throughput was 10/h. The immobilized enzyme reactor was stable for at least 2 months.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.