Abstract
A new system has been developed for rapid on-line SPE-LC-MS-MS analysis of a wide spectrum of trace level pesticides in drinking and surface water. The system employs a serial arrangement of two different turbulent flow columns (TFC; polymer based column in front, carbon based column back) for rapid extraction (Turbulent flow solid phase extraction; TFSPE) from small sample volumes (10 mL) at high flow rates (5 mL min−1). Purely organic eluent is used for TFC elution, and the effluent is re-mixed with water in a mixing chamber in front of the HPLC column to provide efficient analyte focusing. This methodology allows to use a short monolithic LC column for the following fast on-line LC-APCI-MS-MS analysis.