Abstract

A new sensor was developed for simultaneous detection of cadmium (Cd 2+), copper (Cu 2+), and lead (Pb 2+), based on the voltammetric response at a carbon paste electrode modified with carbamoylphosphonic acid (acetamide phosphonic acid) self-assembled monolayer (SAM) on mesoporous silica (Ac-Phos SAMMS). The adsorptive stripping voltammetry (AdSV) technique involves preconcentration of the metal ions onto Ac-Phos SAMMS under an open circuit, then electrolysis of the preconcentrated species, followed by a square wave potential sweep towards positive values. Factors affecting the preconcentration process were investigated. The voltammetric responses increased linearly with the preconcentration time from 1 to 30 min or with metal ion concentrations ranging from 10 to 200 ppb. The responses also evolved in the same fashion as adsorption isotherm in the pH range of 2–6. The metal detection limits were 10 ppb after 2 min preconcentration and improved to 0.5 ppb after 20 min preconcentration.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.